Facilitating Mindfulness Training with Ultrasonic Neuromodulation
Brian Lord, Erica N. Lord, Jessica Schachtner, Laura Beaman, Shinzen Young, John J. B. Allen, Joseph L. Sanguinetti
2026 · 10.64898/2026.03.10.710890
Abstract
Systematic focus training is increasingly recognized for its therapeutic benefits, with equanimity, the ability to maintain an open and accepting attitude towards all experience, identified as a critical mechanism for improving well-being. Physiologically, experienced meditators demonstrate reduced activity in the posterior cingulate cortex (PCC), a hub of the default mode network (DMN), and increased segregation between the DMN and the central executive network (CEN). This study investigated whether non-invasive neuromodulation could facilitate these neural and behavioral shifts in novice practitioners. We conducted a single-blind, randomized controlled trial with 24 meditation-naïve participants who engaged in a two-week “Body Focus” mindfulness training program. Participants were randomized to receive either active (n=16) or sham (n=8) suppressive transcranial focused ultrasound (tFUS) targeting the PCC during four in-person meditation sessions. Resting-state fMRI analysis revealed a robust Condition × Session interaction in functional connectivity. While the sham group showed a trend toward increased coupling, the active tFUS group demonstrated significant decoupling (increased segregation) between the DMN and CEN, a pattern characteristic of advanced meditators. Subnetwork analysis indicated these effects were driven primarily by the decoupling of the core self-referential system in DMN (DMN A ) from the external-oriented control system of the CEN (CEN B ). Behaviorally, greater reductions in DMN-CEN connectivity within the active group predicted larger increases in self-reported acceptance and longer duration of voluntary meditation practice. These findings suggest that tFUS targeting the PCC can acutely redirect neuroplastic trajectories during early mindfulness training, potentially accelerating the acquisition of equanimity and distinct network configurations associated with effortless awareness. Significance Statement This study demonstrates that transcranial focused ultrasound (tFUS) targeting the posterior cingulate cortex (PCC) can synergistically enhance mindfulness training in novices. By inducing a robust decoupling between the default mode and central executive networks – a neural signature typically acquired only after hundreds of hours of practice – this intervention effectively redirected the neural trajectory of novice practitioners toward that of experienced meditators in just two weeks. These findings suggest that targeted neuromodulation can bypass early obstacles in meditation practice, offering a promising “precision wellness” avenue for accelerating the acquisition of equanimity and its associated well-being benefits.
Abstract via europepmc.
Exposures
Exposure 1: Active tFUS to PCC during meditation, four sessions
Target: posterior cingulate cortex — “Peak PCC activity in DMN”
Device: Blatek · Blatek · AT35246 (single-element, 55 mm fixed depth) ✓
| Waveform | pulsed | |
|---|---|---|
| Fundamental frequency (kHz) | 500 | ✓✓✓ |
| Pulse duration (ms) | not reportedimplied by duty cycle ÷ PRF: 0.05 ms (not stated by the paper) | ⚑ |
| Pulse repetition frequency (Hz) | 1,000 | ✓✓✓⚑ |
| Duty cycle (%) | 5 | ✓✓✓ |
| Sonication duration (s) | 5 | ✓✓✓⚑ |
| Free-field pressure (kPa) | not reported | |
|---|---|---|
| Free-field Isppa (W/cm²) | not reported | |
| Free-field Ispta (W/cm²) | not reported | |
| In-situ estimate | not reported | |
| In-situ pressure (kPa) | not reported | |
| In-situ Isppa (W/cm²) | not reported | |
| In-situ Ispta (W/cm²) | not reported | |
| Isppa, domain unspecified (W/cm²) | 3.88 | ✓✓✓⚑ |
| Ispta, domain unspecified (W/cm²) | 1.94 | ✓✓✓⚑ |
5% duty cycle; stimulus pattern 5 seconds on, 10 seconds off for 5 minutes; delivered by hand-held transducer during meditation; PRF 1000 Hz inferred by authors from audible tone.
Flags from extraction
exposures[0].unspecified_domain.ispta_w_cm2— Table gives 1938.61 mW/cm2 (recorded as 1.93861 W/cm2) but ISPPA 3.88 W/cm2 at 5% duty cycle would imply about 194 mW/cm2; possible typo, recorded as stated.exposures[0].timing.pulse_repetition_frequency_hz— PRF appears only in the results as the 'audible 1000 Hz PRF'; not in the parameter table.exposures[0].timing.sonication_duration_s— Recorded the 5 s on period of the 5 s on / 10 s off pattern; total protocol 5 minutes.n_subjects— 24 completed; text says 3 withdrew and 7 eligible did not start, so enrolled number is unclear.exposures[0].timing.pulse_duration_ms— Not stated.exposures[0].unspecified_domain.isppa_w_cm2— Intensity domain not stated; no skull correction applied per limitations.