Quantitative photoacoustic evaluation of graded ischemic stroke and the therapeutic efficacy of low-intensity transcranial ultrasound stimulation

Rui Hu, Li Liu, Zijian Hua, Jingxian Wang, Ying Li, Dean Ta

Biomedical Optics Express 2026 · 10.1364/boe.593436

rodentstrokecerebral haemodynamicshistology molecular

Abstract

Ischemic stroke represents a leading cause of global mortality and long-term disability. Consequently, rapid quantification of cerebral damage severity, coupled with timely neuromodulatory intervention, is imperative for improving clinical prognosis. In this study, we present a label-free quantitative monitoring framework utilizing optical-resolution photoacoustic microscopy (OR-PAM) to assess microvascular alterations and evaluate the therapeutic efficacy of low-intensity transcranial ultrasound stimulation (LITUS). A graded ischemic stroke model was established by modulating photothrombotic irradiation duration (3, 5, and 10 min) to validate system sensitivity. Subsequently, a Hessian filter-based segmentation pipeline was employed to extract quantitative vascular metrics. Five key morphological and functional parameters, including the photoacoustic (PA) signal intensity, vessel area fraction (VAF), average vessel diameter (AVD), branch-point number (BN), and perfused vessel density (PVD), were extracted to characterize the severity of ischemic injury. Our analysis revealed a strong negative correlation between irradiation duration and vascular perfusion-related metrics, which was further confirmed by ex vivo 2,3,5-triphenyltetrazolium chloride (TTC) staining showing duration-dependent infarct expansion. Applying this quantitative framework to the established model, we further demonstrated that acute LITUS treatment significantly alleviated microvascular hypoperfusion and promoted rapid hemodynamic recovery by restoring both functional perfusion and vascular morphology via vasodilation. These findings highlight the potential of a PAM-based quantitative framework for accurate grading of ischemic severity and evaluation of ultrasound-based neuromodulation.

Abstract via europepmc.

Speciesmouse (C57BL/6J)
Subjects5 animals
Sessions per subjectnot reported
Randomisedyes
Blindingsingle
Sham / controlno treatment control
Auditory controlnot reported
Readout timingoffline
Anaesthesiaanaesthetised
Readoutscerebral haemodynamics, histology molecular — Optical-resolution photoacoustic microscopy (PA intensity, vessel area fraction, vessel diameter, branch points, perfused vessel density); TTC infarct staining
Direction of effectnot assessedLITUS after photothrombotic stroke restored PA signal intensity, perfused vessel density and vessel diameter toward baseline and reduced infarct area at 24 h versus untreated PT mice.
Adverse eventsnot reportedCortical temperature elevation below 2 C; mechanical index calculated as 1.16.

Exposures

Exposure 1: LITUS of photothrombotic lesion, left cortex

Target: cerebral cortex — “cortical target coordinates (AP = -2.0 mm, ML = -2.0 mm)”
Device: other named manufacturer · Siansonic · AQ20-1020 ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)1,000✓✓✓
Pulse duration (ms)not reportedimplied by duty cycle ÷ PRF: 0.2 ms (not stated by the paper)⚑
Pulse repetition frequency (Hz)1,000✓✓✓
Duty cycle (%)20✓✓✓
Sonication duration (s)0.3✓✓✓⚑
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatenot reported
In-situ pressure (kPa)not reported
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Pressure, domain unspecified (kPa)1,160✓✓✓⚑
Ispta, domain unspecified (W/cm²)0.45✓✓✓
Protocol, in the paper’s words

PRF 1 kHz, duty cycle 20%, sonication duration 300 ms; total treatment duration 10 min starting 23 min after the photothrombotic procedure.

Flags from extraction

  • n_subjects — 35 mice enrolled across two cohorts; only the PT + LITUS group (five animals) received ultrasound.
  • blinding — Only image acquisition and quantitative analysis stated blinded.
  • exposures[0].timing.pulse_duration_ms — Pulse duration not stated; not derived from duty cycle and PRF.
  • exposures[0].unspecified_domain.pressure_kpa — 1.16 MPa described as measured peak negative pressure; medium not stated.
  • exposures[0].target.terms — Target given only as stereotaxic coordinates AP -2.0, ML -2.0 mm in the left cortex.
  • anaesthesia — Isoflurane stated for stroke induction; anaesthesia during LITUS not stated explicitly.
  • exposures[0].timing.sonication_duration_s — 300 ms sonication within a 10 min total treatment; inter-sonication interval not stated.