Brain injury during focused ultrasound neuromodulation for substance use disorder
Ali Rezai, Manish Ranjan, Aniruddha Bhagwat, Tasneem Arsiwala, Jeffrey Carpenter, Mark Schafer, Geoffrey Adams, Jennifer Marton, Padmashree Tirumalai, Daniel Farmer, James Mahoney, Victor Finomore
Brain Stimulation 2025, 18, 2050-2053 · 10.1016/j.brs.2025.10.012
Exposures
Exposure 1: Bilateral NAc FUS neuromodulation (active treatment arm, adverse event case)
Target: nucleus accumbens — “bilateral nucleus accumbens (NAc)”
Device: Insightec · Insightec · Insightec 220kHz system (Exablate Neuro) ✓
| Waveform | pulsed | |
|---|---|---|
| Fundamental frequency (kHz) | 220 | ✓✓✓ |
| Pulse duration (ms) | 100 | ✓✓✓ |
| Pulse repetition frequency (Hz) | 0.333 | ✓✓✓⚑ |
| Duty cycle (%) | 3.3pulse duration × PRF gives 3.33% | ✓✓✓ |
| Sonication duration (s) | 300 | ✓✓✓ |
| Free-field pressure (kPa) | not reported | |
|---|---|---|
| Free-field Isppa (W/cm²) | not reported | |
| Free-field Ispta (W/cm²) | not reported | |
| In-situ estimate | not reported | |
| In-situ pressure (kPa) | not reported | |
| In-situ Isppa (W/cm²) | not reported | |
| In-situ Ispta (W/cm²) | not reported |
Each FUS treatment block consisted of bilateral NAc sonication lasting 300 seconds (up to 6 blocks), delivering therapeutic energy up to 100 W, with four individual sub-spots per NAc target treated sequentially at an effective pulse repetition time of 3 seconds. In the reported adverse event, an initial 90 W sonication was halted at 89 seconds by the acoustic feedback safety system; a second 90 W sonication then ran the full 300 seconds without a halt, after which the participant became unarousable.
Flags from extraction
exposures[0].free_field— Paper explicitly states pressure and intensity values are proprietary to the manufacturer and not disclosed: 'Technical information regarding pressure intensities and mechanical index are unique and proprietary to Insightec.'sham_type— The mechanism of the sham condition is not described in this letter; recorded as 'other'.exposures[0].timing.pulse_repetition_frequency_hz— PRF derived by converting the stated effective pulse repetition time (3 seconds) to Hz (1/3 = 0.333 Hz), consistent with the stated 3.3% duty cycle and 100 ms pulse duration.