A pilot safety and tolerability study of scanning ultrasound as a neuromodulation therapy in Alzheimer’s disease

Peter J Nestor, Matthew Pelekanos, Gerhard Leinenga, Jae Song, Wendy Lee, Gina Richter-Stretton, Caitlin McElligott, Amir Fazlollahi, Jason B Mattingley, Anthony Harris, Henry Beale, Jennie Roberts, Rachel de las Heras, Jürgen Götz

Brain Communications 2025 · 10.1093/braincomms/fcaf445

human patientalzheimers diseaseclinical scaleeeg megfmriother mri

Abstract

Clearing amyloid-β pathology in Alzheimer's disease (AD) has been considered a prerequisite for restoring cognitive functions. Intriguingly, by application of a modality of scanning ultrasound (SUS) to mice that does not remove amyloid-β, we previously achieved significant cognitive improvements. This prompted us to explore SUS as a non-invasive brain stimulation strategy in an open-label safety trial in AD. We conducted a human pilot study in 12 participants with AD with the primary objective of determining feasibility, safety and tolerability. Exploratory secondary end-points were cognitive and behavioural measures, resting-state EEG and functional MRI. A portable device termed UltraThera Pilot was built under medical device standard guidelines, integrating a Brainsight image-guided neuronavigation system. A single-element 286-kHz transducer was programmed to deliver non-derated ultrasound doses of 2.6, 1.95 or 1.3 MPa. With four treatment sessions spaced fortnightly, four participants received 30 sonications per session (precuneus, ∼30 cm 3 brain tissue) and the remaining 8 received 100 sonications per session (bilateral precuneus and temporo-parietal association cortex, ∼100 cm 3 ). Safety monitoring, EEG, MRI, cognitive and neuropsychiatric evaluations were performed. The treatment was fast, safe and well-tolerated at the 1.95 MPa dose. MRI showed no changes, whereas changes were observed in aperiodic EEG content. Cognitive performance did not change but statistically significant improvements in behavioural and psychological symptoms were found using the Neuropsychiatric Inventory test. In conclusion, this SUS safety trial met its primary and secondary end-points in biomarker-confirmed mild-to-moderate AD. It informs our future work in an upcoming efficacy trial in an AD population.

Abstract via europepmc.

Specieshuman
Subjects12 participants
Sessions per subject4
Randomisedno
Blindingnone
Sham / controlnone
Auditory controlpost hoc check
Readout timingoffline
Anaesthesianot applicable
Readoutsclinical scale, eeg meg, fmri, other mri — ACE-III; ADAS-Cog; 12-item NPI; resting-state EEG (aperiodic slope, alpha); resting-state fMRI; diffusion MRI
Direction of effectmixed or unclearOpen-label: NPI scores improved and EEG aperiodic slope steepened after treatment; cognition (ACE-III, ADAS-Cog), MRI and standard EEG measures did not change.
Adverse eventsobservedOne treatment-emergent adverse event related to the procedure (scalp irritation from head shaving and wig); one unrelated serious adverse event; the 2.6 MPa dose caused pain with flinching in all four Cohort 1 participants and was abandoned; participants sometimes heard sound.

Exposures

Exposure 1: SUS to bilateral precuneus/posterior cingulate

Target: precuneus, posterior cingulate cortex — “bilateral precuneus/posterior cingulate region”
Device: Sonic Concepts · Sonic Concept Inc., USA · H117 ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)286✓✓✓
Pulse duration (ms)10✓✓✓
Pulse repetition frequency (Hz)10✓✓✓
Duty cycle (%)not reportedpulse duration × PRF gives 10%⚑
Sonication duration (s)6✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)1,950, 2,600swept✓✓✓⚑
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatederatingsingle value
In-situ pressure (kPa)780, 1,040swept✓✓✓⚑
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Isppa, domain unspecified (W/cm²)21.1, 37.5swept✓✓✓⚑
Protocol, in the paper’s words

Scanning mode: 6-s sonications (10 ms pulses, 10 Hz PRF) moved across a 6 x 6 mm target grid, 30 sonications per session in Cohort 1 and 100 in Cohort 2; four sessions fortnightly. Started at 2.6 MPa free-field PNP, titrated to 1.95 MPa (all of Cohort 2 at 1.95 MPa).

Exposure 2: SUS to bilateral lateral temporo-parietal association cortex (Cohort 2 only)

Target: temporoparietal junction — “bilateral temporo-parietal association cortex”
Device: Sonic Concepts · Sonic Concept Inc., USA · H117 ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)286✓✓✓
Pulse duration (ms)10✓✓✓
Pulse repetition frequency (Hz)10✓✓✓
Duty cycle (%)not reportedpulse duration × PRF gives 10%
Sonication duration (s)6✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)1,950✓✓✓
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatederatingsingle value
In-situ pressure (kPa)780✓✓✓
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Isppa, domain unspecified (W/cm²)21.1✓✓✓
Protocol, in the paper’s words

Scanning mode: 6-s sonications (10 ms pulses, 10 Hz PRF) moved across a 6 x 6 mm target grid, 30 sonications per session in Cohort 1 and 100 in Cohort 2; four sessions fortnightly. Started at 2.6 MPa free-field PNP, titrated to 1.95 MPa (all of Cohort 2 at 1.95 MPa).

Flags from extraction

  • exposures[0].free_field.pressure_kpa — Peak negative pressure in free field for three dose levels (100%, 75%, 50%); 1.95 MPa was the dose delivered for nearly all sonications; 2.6 MPa not tolerated.
  • exposures[0].in_situ.pressure_kpa — Derated PNP values are upper bounds ('<=') based on an assumed 64% average skull attenuation measured in sheep skulls, not subject-specific.
  • exposures[0].unspecified_domain.isppa_w_cm2 — Table 2 Isppa values (<=37.5, <=21.1, <=9.4 W/cm2) are not explicitly assigned to free field or derated conditions; discussion pairs 21.1 W/cm2 with 'external PNP of 1.95 MPa'.
  • exposures[0].timing.duty_cycle_pct — Duty cycle not stated; not computed.
  • exposures[1].target.terms — Mapped 'temporo-parietal association cortex' to temporoparietal_junction as nearest vocabulary term.
  • auditory_control — Paper reports participants sometimes heard sound; no auditory control was used or described.