Yongxu Ju, Xiao Wang, Yang Liu, Chuan Chen, Fang Yang

ACS Sensors 2025 · 10.1021/acssensors.5c00258

in vitro cellepilepsyinvasive electrophysiologycellular imaginghistology molecular

Abstract

Ultrasound neuromodulation holds great potential for treating neurological disorders. However, the molecular mechanisms underlying its effects remain unclear. There is a lack of systematic safety evaluation and standardized quantification of therapeutic outcomes. To address these challenges, this study integrates on-chip ultrasound with multiparameter cellular monitoring to quantitatively assess the therapeutic effects of various ultrasound parameters on neurological diseases from both electrophysiological and metabolic perspectives. Long-term monitoring is employed for a comprehensive evaluation of ultrasound safety. Notably, the combination of high-resolution, synchronous signal monitoring and biochemical indicators provides a deeper and more thorough understanding of the molecular mechanisms involved in ultrasound neuromodulation. Furthermore, using an on-chip epileptic neuron model, it is demonstrated that epileptic neurons exhibit high-frequency firing and synchronized oscillations in excitability, temperature, intracellular calcium, and mitochondrial calcium concentrations. The application of on-chip ultrasound effectively suppresses epileptiform discharges in neurons and inhibits both these oscillations and neuronal bursting events. When tuning with acoustic enhancer sodium valproate-loaded nanobubbles, the antiepileptic therapy effect can be significantly enhanced by maintaining mitochondrial calcium homeostasis. By integrating multiparameter long-term monitoring with on-chip ultrasound, this study presents an advanced approach for investigating the mechanisms of ultrasound neuromodulation, facilitating standardized evaluation and long-term safety assessment.

Abstract via europepmc.

Speciesnot reported
Subjectsnot reported cultures
Sessions per subjectnot applicable
Randomisednot applicable
Blindingnot applicable
Sham / controlno treatment control
Auditory controlnot reported
Readout timingboth
Anaesthesianot applicable
Readoutsinvasive electrophysiology, cellular imaging, histology molecular — on-chip microelectrode firing frequency, platinum resistor temperature, Ca2+ and ROS imaging, viability
Direction of effectinhibitoryUS1 and US2 reduced epileptic neuron firing frequency and temperature oscillation; US3 and US4 had no therapeutic effect.
Adverse eventsnot applicableLIPUS parameters US1-US4 showed no apoptosis, necrosis or cytoskeletal disruption in normal neurons.

Exposures

Exposure 1: On-chip LIPUS (US1-US4) on epileptic cultured neurons

Target: cultured neurons — “on-chip epileptic neurons”
Device: custom-built · interdigital transducer (IDT) microelectrode and platinum resistor array chip

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)1,000✓✓✓
Pulse duration (ms)not reported
Pulse repetition frequency (Hz)not reported
Duty cycle (%)30✓✓✓
Sonication duration (s)300✓✓✓⚑
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatenot applicable
In-situ pressure (kPa)not applicable
In-situ Isppa (W/cm²)not applicable
In-situ Ispta (W/cm²)not applicable
Pressure, domain unspecified (kPa)30✓?⚑
Protocol, in the paper’s words

Four LIPUS parameter sets (US1-US4, listed in supplementary Table S1) were applied to Mg2+-free epileptic neuron cultures on a surface acoustic wave chip; 1 MHz, 30 kPa and 30% duty cycle is the parameter set the abstract reports as suppressing discharges; US2 was applied for 5 min in the heating test.

Flags from extraction

  • species — Species of the cultured neurons not stated in the main text.
  • exposures[0].unspecified_domain.pressure_kpa — Abstract gives 1 MHz, 30 kPa, 30% duty cycle; full US1-US4 parameters are in a supplementary table not in the text.
  • exposures[0].timing.sonication_duration_s — Only a 5 min application for US2 in the heating test is mentioned; protocol otherwise in supplement.