Disrupting action control with transcranial ultrasound neuromodulation: a step forward for Tourette syndrome

Cyril Atkinson-Clement, Mairi Houlgreave, Aikaterini Gialopsou, Caitlin Mairi Smith, Isabel Farr, Aneta Dvorakova, James Kennaway, Stephen R. Jackson

2025 · 10.1101/2025.10.29.685348

human healthyhealthybehaviour

Abstract

Background While Tourette syndrome (TS) is characterised by the presence of tics, premonitory urges are involved in both triggering them and as a lever to control them. The right insular cortex is known to play a key role in these processes, yet its precise functional contribution remains unclear. Transcranial ultrasound stimulation (TUS) is currently one of the only non-invasive neuromodulation techniques capable of safely and precisely modulating the activity of deep brain structures such as the insula. Objective This study used TUS to causally investigate the respective roles of the anterior and posterior insula in urge and action control during a blink suppression task. Methods Twenty healthy adults underwent three TUS sessions targeting the right anterior insula, posterior insula, and posterior ventricle (control site) in a within-subject, crossover design. Behavioural outcomes included blink counts and urge ratings. Additionally, voxel-based correlations were conducted to further estimate the role of each reached area with acoustic stimulation, allowing for a better mechanical understanding. Results TUS of the anterior insula induced a clear and acute increase in blink frequency without altering urge, reinforced by voxel-based analyses, confirming its role in voluntary action control. Posterior insula stimulation induced delayed increases in both blinks and urge, which are essentially explained by an involuntary stimulation of the right inferior frontal gyrus. Ventricular stimulation unexpectedly reduced blink frequency during the stimulation only, correlated with acoustic energy delivered within the posterior internal capsule. No significant side effects or temperature rises were observed. Conclusions These findings provide the first causal evidence that TUS can modulate the two main features of TS: premonitory urge and action control. Distinct effects across targets highlight the anterior insula’s critical role in action control and suggest the internal capsule as a potential therapeutic target for TS.

Abstract via europepmc.

Specieshuman
Subjects20 participants
Sessions per subject3
Randomisednot reported
Blindingnot reported
Sham / controlactive control site
Auditory controlmasking sound
Readout timingboth
Anaesthesianot applicable
Readoutsbehaviour — blink counts from EOG; urge-to-blink ratings; TUS detection ratings
Direction of effectmixed or unclearAnterior insula TUS acutely increased blink frequency without altering urge; posterior insula TUS produced delayed increases in blinks and urge; ventricle (control site) TUS reduced blink frequency during stimulation only.
Adverse eventsnone observedSide-effect questionnaire before, immediately after and 24 h after TUS; scores were higher before TUS than after; no temperature rises or significant side effects reported.

Exposures

Exposure 1: TUS-AntIns

Target: anterior insula — “right anterior insula”
Device: NeuroFUS · Sonic Concepts (sold by Brainbox) · CTX-500-4CH transducer with NeuroFUS PRO TPO-203 ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)500✓✓✓
Pulse duration (ms)20✓✓✓
Pulse repetition frequency (Hz)5✓✓✓
Duty cycle (%)10pulse duration × PRF gives 10%✓✓✓
Sonication duration (s)80✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)54.51✓✓✓
Free-field Ispta (W/cm²)not reported
In-situ estimatesimulationmean or range across subjects
In-situ pressure (kPa)386.4✓✓✓⚑
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Protocol, in the paper’s words

Single 80 s sonication delivered during block 5 of a 15-block (3 min each) blink suppression task; pulse duration 20 ms, pulse repetition interval 200 ms. Identical parameters at all three targets; one target per visit, three visits at least one week apart.

Exposure 2: TUS-PostIns

Target: posterior insula — “right posterior insula”
Device: NeuroFUS · Sonic Concepts (sold by Brainbox) · CTX-500-4CH transducer with NeuroFUS PRO TPO-203 ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)500✓✓✓
Pulse duration (ms)20✓✓✓
Pulse repetition frequency (Hz)5✓✓✓
Duty cycle (%)10pulse duration × PRF gives 10%✓✓✓
Sonication duration (s)80✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)54.51✓✓✓
Free-field Ispta (W/cm²)not reported
In-situ estimatesimulationmean or range across subjects
In-situ pressure (kPa)476.4✓✓✓⚑
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Protocol, in the paper’s words

Single 80 s sonication delivered during block 5 of a 15-block (3 min each) blink suppression task; pulse duration 20 ms, pulse repetition interval 200 ms. Identical parameters at all three targets; one target per visit, three visits at least one week apart.

Exposure 3: TUS-Vent

Target: other — “right posterior lateral ventricle”
Device: NeuroFUS · Sonic Concepts (sold by Brainbox) · CTX-500-4CH transducer with NeuroFUS PRO TPO-203 ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)500✓✓✓
Pulse duration (ms)20✓✓✓
Pulse repetition frequency (Hz)5✓✓✓
Duty cycle (%)10pulse duration × PRF gives 10%✓✓✓
Sonication duration (s)80✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)54.51✓✓✓
Free-field Ispta (W/cm²)not reported
In-situ estimatesimulationmean or range across subjects
In-situ pressure (kPa)546.8✓✓✓
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Protocol, in the paper’s words

Single 80 s sonication delivered during block 5 of a 15-block (3 min each) blink suppression task; pulse duration 20 ms, pulse repetition interval 200 ms. Identical parameters at all three targets; one target per visit, three visits at least one week apart.

Flags from extraction

  • exposures[1].in_situ.pressure_kpa — Text gives 476.36 kPa for TUS-PostIns but Table 1 mean row gives 479.4; text value recorded. Text gives 546.77 and 386.35 versus table 546.8 and 386.3.
  • exposures[0].in_situ.pressure_kpa — Peak pressure is a k-Plan simulated value in brain tissue, mean across 20 participants (SD in quote).
  • exposures[2].target.terms — Posterior lateral ventricle used as control site; no suitable target id so 'other'.
  • sham_type — No sham; the ventricle target is a real-ultrasound control site.
  • auditory_control — Bone-conduction headphones played a sound imitating TUS plus white noise; participants still detected stimulation above chance in some blocks.