Assessment of focused ultrasound stimulation to induce peripheral nerve activity and potential damage in vivo
Bruno Rodríguez-Meana, Eva Santos-Nogueira, Sònia Trujillo-Vázquez, Anette Jakob, Esther Udina, Marc Fournelle, Xavier Navarro
Frontiers in Neurology 2024 · 10.3389/fneur.2024.1346412
Abstract
Introduction Peripheral neuroprostheses are aimed to restore loss of sensory and motor functions by interfacing axons in the peripheral nerves. Most common interfaces in neuroprostheses are electrodes that establish electrical connection with peripheral axons. However, some challenges arise related to long-term functionality, durability, and body response. Recently, focused ultrasound stimulation (FUS) has emerged as a non-invasive approach to modulate the nervous system. However, it is controversial whether FUS can induce axon depolarization. Methods We have assessed FUS applied in vivo to the rat peripheral nerve, with two objectives: first, to test whether FUS activates peripheral nerves under different stimulation conditions, and second, to evaluate if FUS inflicts damage to the nerve. FUS was delivered with three ultrasound transducers (Sonic Concept H115, H107, and H102) covering the largest set of parameters examined for FUS of peripheral nerves so far. Results We did not obtain reliable evoked action potentials in either nerves or muscles, under any FUS condition applied, neither over the skin nor directly to the nerve exposed. Additional experiments ex vivo and in vivo on mice, confirmed this conclusion. When FUS stimulation was applied directly to the exposed sciatic nerve, neuromuscular function decreased significantly, and recovered one week later, except for FUS at 0.25 MHz. Histologically, degenerating nerve fibers were observed, with a tendency to be higher with the lower FUS frequency. Discussion Past reports on the ability of ultrasound to stimulate the peripheral nerve are controversial. After testing a wide range of FUS conditions, we conclude that it is not a reliable and safe method for stimulating the peripheral nerve. Special consideration should be taken, especially when low-frequency FUS is applied, as it may lead to nerve damage.
Abstract via europepmc.
Exposures
Exposure 1: Sciatic nerve (through skin, exposed nerve) and hindpaw, five transducer frequencies
Target: sciatic nerve — “sciatic nerve”
Device: Sonic Concepts · Sonic Concepts · H115, H107, H102 (H107 and H102 also at 3rd harmonic)
| Waveform | pulsed | |
|---|---|---|
| Fundamental frequency (kHz) | 250, 500, 1,120, 1,630, 3,580swept | ✓✓✓⚑ |
| Pulse duration (ms) | 1, 10, 50, 100, 200swept | ✓✓✓ |
| Pulse repetition frequency (Hz) | not reported | ⚑ |
| Duty cycle (%) | not reported | |
| Sonication duration (s) | not reported |
| Free-field pressure (kPa) | not reported | |
|---|---|---|
| Free-field Isppa (W/cm²) | not reported | |
| Free-field Ispta (W/cm²) | not reported | |
| In-situ estimate | not reported | |
| In-situ pressure (kPa) | not reported | |
| In-situ Isppa (W/cm²) | not reported | |
| In-situ Ispta (W/cm²) | not reported | |
| Pressure, domain unspecified (kPa) | 100, 5,000swept | ✓?⚑ |
5 sequences of 1, 3 or 10 pulses of 1, 10, 50, 100 or 200 ms with an interstimulus interval of 100 ms; intersequence interval 1 s; pressure stepped 0, 1.25, 2.5, 3.75 and 5 MPa on each sequence.
Flags from extraction
n_subjects— 22 rats and 4 mice stated; total not stated separately and one ex vivo experiment used a rat that may be among the 22.exposures[0].fundamental_frequency_khz— Five frequencies (H115 0.25, H107 0.5, H102 1.12 MHz and 3rd harmonics 1.63 and 3.58 MHz) recorded as a list in one exposure; focal depth and parameters otherwise shared.exposures[0].unspecified_domain.pressure_kpa— Pressure range 0.1-5 MPa described as 'effective pressure'; transducers were calibrated in water under free-field conditions but the paper does not say the stated pressures are free-field values. Range also read from Table 1.exposures[0].timing.pulse_repetition_frequency_hz— Only an interstimulus interval of 100 ms between pulses is given; PRF, duty cycle and sonication duration are not stated.sham_type— No FUS sham; electrical stimulation used as a control for recording and nerve function.