Closed-loop transcranial ultrasound stimulation with a fuzzy controller for modulation of motor response and neural activity of mice

Yi Yuan, Ai Long, Yongkang Wu, Xiaoli Li

Journal of Neural Engineering 2022 · 10.1088/1741-2552/ac7893

rodenthealthyemg mepinvasive electrophysiology

Abstract

Objective . We propose a closed-loop transcranial ultrasound stimulation (TUS) with a fuzzy controller to realize real-time and precise control of the motor response and neural activity of mice. Approach . The mean absolute value (MAV) of the electromyogram (EMG) and peak value (PV) of the local field potential (LFP) were measured under different ultrasound intensities. A model comprising the characteristics of the MAV of the EMG, PV of the LFP, and ultrasound intensity was built using a neural network, and a fuzzy controller, proportional-integral-derivative (PID) controller, and immune feedback controller were proposed to adjust the ultrasound intensity using the feedback of the EMG MAV and the LFP PV. Main results . In simulation, the quantitative calculation indicated that the maximum relative errors between the simulated EMG MAV and the expected values were 17% (fuzzy controller), 110% (PID control), 66% (immune feedback control); furthermore, the corresponding values of the LFP PV were 12% (fuzzy controller), 53% (PID control), 55% (immune feedback control). The average relative errors of fuzzy controller, PID control, immune feedback control were 4.97%, 13.15%, 11.52%, in the EMG closed-loop experiment and 7.76%, 11.84%, 13.56%, in the LFP closed-loop experiment. Significance . The simulation and experimental results demonstrate that the closed-loop TUS with a fuzzy controller can realize the tracking control of the motor response and neural activity of mice.

Abstract via europepmc.

Speciesmouse (C57BL)
Subjects20 animals
Sessions per subjectnot reported
Randomisednot reported
Blindingnot reported
Sham / controlnone
Auditory controldeafened subjects
Readout timingonline
Anaesthesiaanaesthetised
Readoutsemg mep, invasive electrophysiology — EMG mean absolute value; LFP peak value
Direction of effectexcitatoryEMG mean absolute value and LFP peak value evoked by TUS increased with ultrasound intensity; a fuzzy controller adjusted intensity to track target EMG/LFP values.
Adverse eventsnot reportedMechanical index and temperature rise are given only in supplementary table S1.

Exposures

Exposure 1: closed-loop TUS of mouse motor cortex, 500 kHz

Target: motor cortex — “mouse motor cortex”
Device: Olympus / Panametrics · Olympus · V301-SU ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)500✓✓✓
Pulse duration (ms)not reportedimplied by duty cycle ÷ PRF: 0.5 ms (not stated by the paper)⚑
Pulse repetition frequency (Hz)1,000✓✓✓
Duty cycle (%)50✓✓✓
Sonication duration (s)0.4✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatenot reported
In-situ pressure (kPa)not reported
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Pressure, domain unspecified (kPa)210, 780swept✓✓✓⚑
Protocol, in the paper’s words

Stimulation duration 400 ms, PRF 1 kHz, duty cycle 50%, stimulation interval 10 s; ultrasound intensity adjusted stimulus by stimulus by a fuzzy, PID or immune feedback controller using EMG or LFP feedback.

Flags from extraction

  • exposures[0].timing.pulse_duration_ms — Burst rule: DC 50% / PRF 1 kHz implies 0.5 ms pulses, so the 400 ms 'stimulation duration' is the train; pulse duration not stated.
  • exposures[0].unspecified_domain.pressure_kpa — Pressure range is tied to generator voltage (350-750 mVpp) and described as sound pressure; domain not stated (hydrophone field map in a water tank was also described).
  • n_subjects — 20 mice used for the open-loop and model experiments; 10 of them in closed-loop runs, plus six additional chemically deafened mice.
  • exposures[0].unspecified_domain.isppa_w_cm2 — Isppa and Ispta are given only in supplementary table S1, not in the main text.
  • sham_type — No sham or control sonication arm; controllers (fuzzy, PID, immune) are compared.