Focused ultrasound neuromodulation on a multiwell MEA

Marta Saccher, Shinnosuke Kawasaki, Martina Proietti Onori, Geeske M. van Woerden, Vasiliki Giagka, Ronald Dekker

Bioelectronic Medicine 2022 · 10.1186/s42234-021-00083-7

in vitro cellhealthyinvasive electrophysiology

Abstract

Background Microelectrode arrays (MEA) enable the measurement and stimulation of the electrical activity of cultured cells. The integration of other neuromodulation methods will significantly enhance the application range of MEAs to study their effects on neurons. A neuromodulation method that is recently gaining more attention is focused ultrasound neuromodulation (FUS), which has the potential to treat neurological disorders reversibly and precisely. Methods In this work, we present the integration of a focused ultrasound delivery system with a multiwell MEA plate. Results The ultrasound delivery system was characterised by ultrasound pressure measurements, and the integration with the MEA plate was modelled with finite-element simulations of acoustic field parameters. The results of the simulations were validated with experimental visualisation of the ultrasound field with Schlieren imaging. In addition, the system was tested on a murine primary hippocampal neuron culture, showing that ultrasound can influence the activity of the neurons. Conclusions Our system was demonstrated to be suitable for studying the effect of focused ultrasound on neuronal cultures. The system allows reproducible experiments across the wells due to its robustness and simplicity of operation.

Abstract via europepmc.

Speciesmouse (FvB/NHsD)
Subjects7 cultures
Sessions per subjectnot applicable
Randomisednot applicable
Blindingnot applicable
Sham / controlnone
Auditory controlnot reported
Readout timingboth
Anaesthesianot applicable
Readoutsinvasive electrophysiology — multi-electrode array extracellular spike and burst recording
Direction of effectmixed or unclearFiring rate increased during ultrasound-on, followed by a reduction or pause in firing, then return to baseline; the response differed between wells.
Adverse eventsnot applicableThermocouple measurement in a water-filled well gave a maximum temperature increase of 0.8 degC over 60 s.

Exposures

Exposure 1: 5 MHz continuous wave, 15 s and 60 s exposures

Target: cultured neurons — “murine primary hippocampal neuron culture”
Device: Olympus / Panametrics · Olympus Panametrics NTD · V310 ✓

Pulse timing
Waveformcontinuous
Fundamental frequency (kHz)5,000✓✓✓
Pulse duration (ms)not applicable
Pulse repetition frequency (Hz)not applicable
Duty cycle (%)not applicable⚑
Sonication duration (s)15, 60swept✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatenot applicable
In-situ pressure (kPa)not applicable⚑
In-situ Isppa (W/cm²)not applicable
In-situ Ispta (W/cm²)not applicable
Protocol, in the paper’s words

Continuous ultrasound at 5 MHz for 15 s or 60 s; each treatment repeated three times with a 5 minute observation period (post period) between treatments; the experiment was performed twice for each protocol.

Flags from extraction

  • exposures[0].in_situ.pressure_kpa — Simulated (COMSOL) peak pressure at the neuron layer in a dish, including standing-wave reflections from the quartz well bottom; there is no skull or tissue path, so the in_situ block may be better read as the at-target value for a dish. Free-field hydrophone pressure was measured but no value is given in the text.
  • n_subjects — Seven MEA wells; the text does not give the number of cultures or animals. Wells treated as preparations.
  • exposures[0].timing.duty_cycle_pct — Continuous wave; duty cycle not stated.