Application of a sub–0.1-mm 3 implantable mote for in vivo real-time wireless temperature sensing
Chen Shi, Victoria Andino-Pavlovsky, Stephen A. Lee, Tiago Costa, Jeffrey Elloian, Elisa E. Konofagou, Kenneth L. Shepard
Science Advances 2021 · 10.1126/sciadv.abf6312
Abstract
There has been increasing interest in wireless, miniaturized implantable medical devices for in vivo and in situ physiological monitoring. Here, we present such an implant that uses a conventional ultrasound imager for wireless powering and data communication and acts as a probe for real-time temperature sensing, including the monitoring of body temperature and temperature changes resulting from therapeutic application of ultrasound. The sub-0.1-mm 3 , sub-1-nW device, referred to as a mote, achieves aggressive miniaturization through the monolithic integration of a custom low-power temperature sensor chip with a microscale piezoelectric transducer fabricated on top of the chip. The small displaced volume of these motes allows them to be implanted or injected using minimally invasive techniques with improved biocompatibility. We demonstrate their sensing functionality in vivo for an ultrasound neurostimulation procedure in mice. Our motes have the potential to be adapted to the distributed and localized sensing of other clinically relevant physiological parameters.
Abstract via europepmc.
Exposures
Exposure 1: FUS of sciatic nerve, mouse hindlimb
Target: sciatic nerve — “sciatic nerves inside the lower hindlimb”
Device: Sonic Concepts · SonicConcepts (Bothell, WA) · SU-107 ✓
| Waveform | pulsed | |
|---|---|---|
| Fundamental frequency (kHz) | 3,570 | ✓✓✓ |
| Pulse duration (ms) | 1 | ✓✓✓⚑ |
| Pulse repetition frequency (Hz) | 10 | ✓✓✓ |
| Duty cycle (%) | not reportedpulse duration × PRF gives 1% | |
| Sonication duration (s) | not reported | ⚑ |
| Free-field pressure (kPa) | not reported | |
|---|---|---|
| Free-field Isppa (W/cm²) | not reported | |
| Free-field Ispta (W/cm²) | not reported | |
| In-situ estimate | not reported | |
| In-situ pressure (kPa) | not reported | |
| In-situ Isppa (W/cm²) | not reported | |
| In-situ Ispta (W/cm²) | not reported | |
| Ispta, domain unspecified (W/cm²) | 0.958, 1.35, 1.61, 1.75swept | ✓✓✓⚑ |
Ultrasonic pulse trains of 1-ms pulse duration and 10-Hz PRF applied to the sciatic nerve through a water-filled coupling cone at four Ispta levels (957.99, 1348.02, 1608.21, 1745.83 mW/cm2); train duration and duty cycle not stated.
Flags from extraction
exposures[0].unspecified_domain.ispta_w_cm2— Four intensities given in mW/cm2 and converted to W/cm2; recorded as [min, max] of the four listed; domain not statedexposures[0].timing.sonication_duration_s— Train duration and duty cycle not stated for the sciatic FUSn_subjects— Number of mice used not stated; motes also used in mice and tissue for non-stimulation sensing experiments (L12-3v imaging probe, 8.3 MHz, not neuromodulation, not recorded as an exposure)exposures[0].timing.pulse_duration_ms— The pulse-train sentence is split across a page break in the extracted text; quoted the continuation