A polydimethylsiloxane-based axicon lens for focused ultrasonic brain stimulation techniques

Fabin Acquaticci, Juan F. Guarracino, Sergio N. Gwirc, Sergio E. Lew

Acoustical Science and Technology 2019 · 10.1250/ast.40.116

rodenthealthybehaviour
Speciesmouse (CF-1)
Subjects3 animals
Sessions per subjectnot reported
Randomisednot reported
Blindingnot reported
Sham / controlnone
Auditory controlnot reported
Readout timingonline
Anaesthesiaanaesthetised
Readoutsbehaviour — Visually observed evoked muscle movements (tail, hindlimb, forelimb)
Direction of effectexcitatoryAxicon-lens tFUS of the mouse motor cortex evoked movement in a specific body segment (tail at the mapped site); US intensity was raised until a movement threshold was seen.
Adverse eventsnone observedNormal behaviour observed in all animals after sonication; intensity stated to be below FDA imaging limits.

Exposures

Exposure 1: Axicon-lens tFUS of mouse motor cortex

Target: motor cortex — “motor cortex (area related to the control of tail motility)”
Device: custom-built · Prototype built in-house with a Steminc piezo-disc element · tFUS-axicon transducer, 445 kHz, epoxy/PDMS axicon lens ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)445✓✓✓
Pulse duration (ms)0.36✓✓✓
Pulse repetition frequency (Hz)1,000✓✓✓
Duty cycle (%)not reportedpulse duration × PRF gives 36%
Sonication duration (s)not reported⚑
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)0.036✓✓✓
In-situ estimatenot reported
In-situ pressure (kPa)not reported
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Pressure, domain unspecified (kPa)100✓✓✓⚑
Protocol, in the paper’s words

Motor cortex stimulated every 3 s with a train of low-intensity tFUS pulses (PD 0.36 ms, PRF 1 kHz); intensity was gradually increased at each site until movement was detected. Number of pulses per train (stimulus duration) not stated.

Flags from extraction

  • exposures[0].unspecified_domain.ispta_w_cm2 — 36 mW/cm2 converted to 0.036 W/cm2; domain not stated (radiant-force power meter; skull effect ignored); also used for mapping at threshold, so may not apply to all conditions.
  • exposures[0].unspecified_domain.pressure_kpa — 0.1 MPa given as 'small peak pressure amplitude' in the discussion; domain not stated and not obviously consistent with 36 mW/cm2 Ispta.
  • exposures[0].timing.sonication_duration_s — Stimulus duration and number of pulses are not given.
  • exposures[0].device.manufacturer — Transducer built in-house; manufacturer field describes the build, piezo element from Steminc.