← Explore

Low-intensity focused ultrasound alters the latency and spatial patterns of sensory-evoked cortical responses in vivo

Jonathan A N Fisher, Iryna Gumenchuk

Journal of Neural Engineering 2018 · 10.1088/1741-2552/aaaee1

rodenthealthycellular imaginghistology molecular

Abstract

Objective The use of transcranial, low intensity focused ultrasound (FUS) is an emerging neuromodulation technology that shows promise for both therapeutic and research applications. Among many, one of the most exciting applications is the use of FUS to rehabilitate or augment human sensory capabilities. While there is compelling empirical evidence demonstrating this capability, basic questions regarding the spatiotemporal extent of the modulatory effects remain. Our objective was to assess the basic, yet often overlooked hypothesis that FUS in fact alters sensory-evoked neural activity within the region of the cerebral cortex at the beam's focus. Approach To address this knowledge gap, we developed an approach to optically interrogate patterns of neural activity in the cortex directly at the acoustic focus, in vivo. Implementing simultaneous wide-field optical imaging and FUS stimulation in mice, our experiments probed somatosensory-evoked electrical activity through the use of voltage sensitive dyes (VSDs) and, in transgenic mice expressing GCaMP6f, monitored associated Ca 2+ responses. Main results Our results demonstrate that low-intensity FUS alters both the kinetics and spatial patterns of neural activity in primary somatosensory cortex at the acoustic focus. When preceded by 1 s of pulsed ultrasound at intensities below 1 W cm -2 (I sppa ), the onset of sensory-evoked cortical responses occurred 3.0 ± 0.7 ms earlier and altered the surface spatial morphology of Ca 2+ responses. Significance These findings support the heretofore unconfirmed assumption that FUS-induced sensory modulation reflects, at least in part, altered reactivity in primary sensory cortex at the site of sonication. The findings are significant given the interest in using FUS to target and alter spatial aspects of sensory receptive fields on the cerebral cortex.

Abstract via europepmc.

Speciesmouse (C57BL/6J; Thy1-GCaMP6f)
Subjects7, 4, 6, 2, 2swept animals
Sessions per subject1
Randomisednot reported
Blindingnot reported
Sham / controlnone
Auditory controlnot reported
Readout timingoffline
Anaesthesiaanaesthetised
Readoutscellular imaging, histology molecular — voltage-sensitive dye and GCaMP6f wide-field imaging; GFAP immunostaining; Evans blue extravasation
Direction of effectmixed or unclearFUS preceding forelimb stimulation shortened onset latency by 3.0 ms at 0.69 W/cm2 (not at 3.5 W/cm2) and increased solidity and circularity of the Ca2+ response; FUS alone evoked no response.
Adverse eventsnone observedNo differential GFAP expression 24 h after sonication and no Evans blue extravasation at the neuromodulation intensity.

Exposures

Exposure 1: FUS to mouse forelimb S1, 510 kHz

Target: primary somatosensory cortex — “forelimb's representation on primary somatosensory cortex”
Device: Sonic Concepts · Sonic Concepts · custom 16-element ring array (modified H-205B), TPO-106 driver ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)510✓✓✓
Pulse duration (ms)0.5✓✓✓
Pulse repetition frequency (Hz)1,000✓✓✓
Duty cycle (%)not reportedpulse duration × PRF gives 50%⚑
Sonication duration (s)1✓✓✓
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatenot reported
In-situ pressure (kPa)not reported
In-situ Isppa (W/cm²)not reported
In-situ Ispta (W/cm²)not reported
Pressure, domain unspecified (kPa)170✓✓✓⚑
Isppa, domain unspecified (W/cm²)0.69, 3.5swept✓✓✓
Protocol, in the paper’s words

A 1 s burst of pulsed FUS (500 us pulses at 510 kHz, 1 kHz repetition rate) ending 200 ms before electrical forelimb stimulation and optical recording of 0.5 s; trials separated by 30 s. Pressure of 0.17 MPa given for the 0.69 W/cm2 condition.

Flags from extraction

  • n_subjects — Group sizes vary by experiment (n=7 and n=4 VSD, n=6 Ca2+, n=2 Evans blue) and no total is given; not_reported.
  • exposures[0].unspecified_domain.pressure_kpa — 0.17 MPa 'peak pressure at the focus' characterised with a hydrophone by the manufacturer; medium not stated, so kept unspecified.
  • exposures[0].timing.duty_cycle_pct — Not stated; 500 us pulses at 1 kHz given.
  • exposures[0].device.family — Custom ring array described as a modified Sonic Concepts H-205B with a Sonic Concepts TPO-106 driver; family assigned by transducer maker.