Transcranial Focused Ultrasound for BOLD fMRI Signal Modulation in Humans
Leo Ai, Jerel K. Mueller, Andrea Grant, Yigitcan Eryaman, Wynn Legon
arXiv 2016 · 10.48550/arxiv.1603.00415
Abstract
Transcranial focused ultrasound (tFUS) is an emerging form of non-surgical human neuromodulation that confers advantages over existing electro and electromagnetic technologies by providing a superior spatial resolution on the millimeter scale as well as the capability to target sub-cortical structures non-invasively. An examination of the pairing of tFUS and blood oxygen level dependent (BOLD) functional MRI (fMRI) in humans is presented here.
Abstract via datacite.
Exposures
Exposure 1: 3T: primary motor cortex hand knob, 0.5 MHz
Target: primary motor cortex — “primary motor cortex hand knob of the dominant hemisphere”
Device: not reported
| Waveform | pulsed | |
|---|---|---|
| Fundamental frequency (kHz) | 500 | ✓✓✓ |
| Pulse duration (ms) | 0.36 | ✓✓✓⚑ |
| Pulse repetition frequency (Hz) | 1,000 | ✓✓✓ |
| Duty cycle (%) | 36pulse duration × PRF gives 36% | ✓✓✓⚑ |
| Sonication duration (s) | 0.5 | ✓✓✓ |
| Free-field pressure (kPa) | not reported | |
|---|---|---|
| Free-field Isppa (W/cm²) | not reported | |
| Free-field Ispta (W/cm²) | not reported | |
| In-situ estimate | measurementsingle value | |
| In-situ pressure (kPa) | not reported | |
| In-situ Isppa (W/cm²) | 6 | ✓✓✓⚑ |
| In-situ Ispta (W/cm²) | not reported |
180 cycles of the 0.5 MHz acoustic frequency were pulsed at 1 KHz (36% duty cycle); 500 cycles of the PRF delivered, a 500 ms stimulus. Event-related design: 90 stimuli of 0.5 s delivered every 6-7 TRs (12-14 s).
Exposure 2: 7T: left head of caudate, 0.86 MHz
Target: caudate nucleus — “left head of caudate”
Device: not reported
| Waveform | pulsed | |
|---|---|---|
| Fundamental frequency (kHz) | 860 | ✓✓✓ |
| Pulse duration (ms) | 0.488 | ✓✓✓ |
| Pulse repetition frequency (Hz) | 1,000 | ✓✓✓ |
| Duty cycle (%) | 50pulse duration × PRF gives 48.8% | ✓✓✓ |
| Sonication duration (s) | 0.5 | ✓✓✓ |
| Free-field pressure (kPa) | not reported | |
|---|---|---|
| Free-field Isppa (W/cm²) | not reported | |
| Free-field Ispta (W/cm²) | not reported | |
| In-situ estimate | measurementsingle value | |
| In-situ pressure (kPa) | not reported | |
| In-situ Isppa (W/cm²) | 6 | ✓✓✓ |
| In-situ Ispta (W/cm²) | not reported |
420 cycles of the 0.86 MHz acoustic frequency (50% duty cycle) pulsed at 1 KHz with a 500 ms duration. Block design, 5 off/on cycles, stimulation delivered every 6 TRs during on cycles.
Flags from extraction
n_subjects— Total never stated; 6 participants at 3T and 1 volunteer at 7T given as group sizes.exposures[0].timing.pulse_duration_ms— Pulse duration not stated in ms; 180 cycles at 0.5 MHz and 420 cycles at 0.86 MHz stated instead. Left not_reported.exposures[0].in_situ.isppa_w_cm2— Isppa of 6 W/cm2 was set after transmission through a 6 mm fragment of human parietal bone in a water tank, not in a brain; recorded as in_situ measurement. Same value stated for both transducers.exposures[0].timing.duty_cycle_pct— Text states 36% for 180 cycles at 0.5 MHz and 1 KHz PRF (arithmetically 18%); recorded as stated.exposures[0].device.family— Transducers not attributed to a maker in the text for the exposures; only tank hydrophone and function generator named.
Notes: Conference-style paper (arXiv 1603.00415); not checked against E8 here.