Image‐guided navigation of single‐element focused ultrasound transducer

Hyungmin Kim, Alan Chiu, Shinsuk Park, Seung‐Schik Yoo

International Journal of Imaging Systems and Technology 2012 · 10.1002/ima.22020

rodenthealthybehaviour

Abstract

The spatial specificity and controllability of focused ultrasound (FUS), in addition to its ability to modify the excitability of neural tissue, allows for the selective and reversible neuromodulation of the brain function, with great potential in neurotherapeutics. Intra-operative magnetic resonance imaging (MRI) guidance (in short, MRg) has limitations due to its complicated examination logistics, such as fixation through skull screws to mount the stereotactic frame, simultaneous sonication in the MRI environment, and restrictions in choosing MR-compatible materials. In order to overcome these limitations, an image-guidance system based on optical tracking and pre-operative imaging data is developed, separating the imaging acquisition for guidance and sonication procedure for treatment. Techniques to define the local coordinates of the focal point of sonication are presented. First, mechanical calibration detects the concentric rotational motion of a rigid-body optical tracker, attached to a straight rod mimicking the sonication path, pivoted at the virtual FUS focus. The spatial error presented in the mechanical calibration was compensated further by MRI-based calibration, which estimates the spatial offset between the navigated focal point and the ground-truth location of the sonication focus obtained from a temperature-sensitive MR sequence. MRI-based calibration offered a significant decrease in spatial errors (1.9±0.8 mm; 57% reduction) compared to the mechanical calibration method alone (4.4±0.9 mm). Using the presented method, pulse-mode FUS was applied to the motor area of the rat brain, and successfully stimulated the motor cortex. The presented techniques can be readily adapted for the transcranial application of FUS to intact human brain.

Abstract via europepmc.

Speciesrat (Sprague-Dawley)
Subjects3 animals
Sessions per subjectnot reported
Randomisednot reported
Blindingnot reported
Sham / controlnone
Auditory controlnot reported
Readout timingonline
Anaesthesiaanaesthetised
Readoutsbehaviour — Tail movement recorded with a piezoelectric pulse transducer
Direction of effectexcitatorySonication of the motor area elicited tail movements synchronised with the 2-s sonication intervals in all animals.
Adverse eventsnot reported

Exposures

Exposure 1: 350 kHz pulsed FUS to rat motor cortex (tail area) with image-guided navigation

Target: motor cortex — “the motor area of the rat brain”
Device: other named manufacturer · SensMed · Sonomo ✓

Pulse timing
Waveformpulsed
Fundamental frequency (kHz)350✓✓✓
Pulse duration (ms)0.5✓✓✓
Pulse repetition frequency (Hz)1,000✓✓✓
Duty cycle (%)50pulse duration × PRF gives 50%✓✓✓
Sonication duration (s)0.3✓✓✓⚑
Pressure and intensity, by domain
Free-field pressure (kPa)not reported
Free-field Isppa (W/cm²)not reported
Free-field Ispta (W/cm²)not reported
In-situ estimatederatingsingle value
In-situ pressure (kPa)500✓✓✓⚑
In-situ Isppa (W/cm²)9.1✓✓✓
In-situ Ispta (W/cm²)4.5✓✓✓
Protocol, in the paper’s words

Pulse-repetition frequency of 1 kHz, tone-burst duration of 0.5 ms, sonication duration of 300 ms and a sonication interval of 2 s, delivered to the site about 2 mm posterior to bregma on the midline.

Flags from extraction

  • exposures[0].in_situ.pressure_kpa — Values stated after applying an assumed 87% rodent skull attenuation from the literature (derating by the authors); free-field values for the in vivo exposure are not given.
  • exposures[0].device.family — SensMed is not a listed family; recorded as other_manufacturer. Quote names the transducer as Sonomo, SensMed.
  • exposures[0].timing.sonication_duration_s — Recorded in ms in the paper (300 ms), converted to 0.3 s.